ICD‐Antiarrhythmic Drug and ICD‐Pacemaker Interactions
- 1 July 1997
- journal article
- review article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 8 (7) , 830-842
- https://doi.org/10.1111/j.1540-8167.1997.tb00842.x
Abstract
Implantable Cardioverter Defibrillators. Antiarrhythmic drugs and separate bradycardia pacing systems are prescribed commonly in patients with implantable cardioverter defibrillators (ICDs). Adverse effects of antiarrhythmic drugs on ICD function and adverse interactions between ICDs and pacemakers have been documented. The effect of antiarrhythmic drugs on the defibrillation threshold (DFT) in patients has not been well assessed. Most studies have been performed in animal models in which cardiac function was normal and drug doses were supraphysiologic. In addition, most studies have utilized monophasic defibrillation shock waveforms and epicardial lead systems. Despite the lack of clinical data applicable to current defibrillation systems, it appears that chronic amiodarone administration causes a significant DFT increase. In addition, antiarrhythmic drugs can influence antitachycardia pacing and tachycardia sensing. Defibrillation shocks can cause transient failure of pacemaker sensing and pacing, and cause spurious pacemaker reprogramming. Pacemaker function can result in ICD oversensing, leading to inappropriate therapy, or cause ICD undersensing, potentially resulting in failure to deliver therapy during ventricular fibrillation. The susceptibility of an ICD to undersensing appears related to the amplitude of the pacing stimulus artifact recorded by the ICD rate‐sensing circuit and to the characteristics of the fibrillation electrogram. Preliminary data suggest that undersensing of ventricular fibrillation by current ICDs is an unlikely event.Keywords
This publication has 91 references indexed in Scilit:
- Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakersAmerican Heart Journal, 1995
- Hemodynamics during ventricular tachyarrhythmiasAmerican Heart Journal, 1994
- Interactions between drugs and devices: Experimental and clinical studiesAmerican Heart Journal, 1994
- Coronary thrombolysis in dogs with intravenously administered human pro-urokinase.Circulation, 1985
- A prospective randomized study of the clinical efficacy and safety of transvenous cardioversion for termination of ventricular tachycardia.Circulation, 1985
- Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator.Circulation, 1985
- Factors that predict syncope during ventricular tachycardia in patientsAmerican Heart Journal, 1984
- Elevation of ventricular defibrillation threshold in dogs by antiarrhythmic drugsAmerican Heart Journal, 1979
- A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing.Circulation, 1977
- Pacemaker Failure Following External DefibrillationCirculation, 1971