Quantitative Analysis of Contribution of Rate Response in Three Different Ventricular Rate Responsive Pacemakers During Out of Hospital Activity
- 1 January 1990
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 13 (1) , 37-44
- https://doi.org/10.1111/j.1540-8159.1990.tb02001.x
Abstract
The contribution of rate response provided by three different types of rate responsive pacemaker (Sensolog, Meta, and Vitatron TX) was assessed in 25 patients during out of hospital activity. Pacemakers were optimally programmed and full disclosure 24-hour Holter ECG recordings were obtained during everyday activities. Close visual analysis of these tracings revealed that rate response is active for less than 15% of the 24-hour study period with high rate response (paced rate above 100 beats/min) for just 5% of the day showing peak activation between 1600 and 2000 hours. Low rate response (paced rate below 100 beats/min but 5 beats above base rate) was maximal between 0800 hours and midday and comprised 10% of all heart beats during the study period. Thirty percent of study patients were nonpacemaker dependent and rate response in these subjects was significantly (P less than 0.001) less active between 0800 hours and midday than in pacemaker dependent patients though it was equivalent at all other times. Nonpacemaker dependent patients achieved natural sinus response in the morning and tended to rely on pacemaker supplied rate response in the latter part of the day.Keywords
This publication has 12 references indexed in Scilit:
- Randomised crossover trial of rate responsive Activitrax and conventional fixed rate ventricular pacing.Heart, 1987
- Rate‐Responsive Pacing by Means of Activity Sensing Versus Single Rate Ventricular Pacing: A Double‐Blind Cross‐Over StudyPacing and Clinical Electrophysiology, 1987
- Rate‐Responsive Pacing: Biosensor Reliability and Physiological SensitivityPacing and Clinical Electrophysiology, 1987
- Increase in Cardiac Output with Rate-Responsive PacemakerThe Annals of Thoracic Surgery, 1986
- Holter-Monitored Heart Rhythm During Atrioventricular Synchronous and Fixed-Rate Ventricular PacingPacing and Clinical Electrophysiology, 1986
- Physiological Versus Single-Rate Ventricular Pacing: A Double-Blind Cross-Over StudyPacing and Clinical Electrophysiology, 1985
- Rate Responsive Pacing Using the TX PacemakerPublished by Springer Nature ,1983
- Arrhythmias documented by 24 hour continuous electrocardiographic monitoring in 50 male medical students without apparent heart diseasePublished by Elsevier ,1977
- Circadian variations of plasma catecholamine, cortisol and immunoreactive insulin concentrations in supine subjectsClinica Chimica Acta; International Journal of Clinical Chemistry, 1974
- Characterization of the Circulatory Response to Maximal Upright Exercise in Normal Subjects and Patients with Heart DiseaseCirculation, 1967