Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device
Open Access
- 1 May 2003
- Vol. 58 (5) , 411-416
- https://doi.org/10.1136/thorax.58.5.411
Abstract
Background: Recent evidence suggests that the frequency response requirements for peak expiratory flow (PEF) meters are higher than was first thought and that the American Thoracic Society (ATS) waveforms to test PEF meters may not be adequate for the purpose. Methods: The dynamic response of mini-Wright (MW), Vitalograph (V), TruZone (TZ), MultiSpiro (MS) and pneumotachograph (PT) flow meters was tested by delivering two differently shaped flow-time profiles from a computer controlled explosive decompression device fitted with a fast response solenoid valve. These profiles matched population 5th and 95th centiles for rise time from 10% to 90% of PEF and dwell time of flow above 90% PEF. Profiles were delivered five times with identical chamber pressure and solenoid aperture at PEF. Any difference in recorded PEF for the two profiles indicates a poor dynamic response. Results: The absolute (% of mean) flow differences in l/min for the V, MW, and PT PEF meters were 25 (4.7), 20 (3.9), and 2 (0.3), respectively, at PEF ≈500 l/min, and 25 (10.5), 20 (8.7) and 6 (3.0) at ≈200 l/min. For TZ and MS meters at ≈500 l/min the differences were 228 (36.1) and 257 (39.2), respectively, and at ≈200 l/min they were 51 (23.9) and 1 (0.5). All the meters met ATS accuracy requirements when tested with their waveforms. Conclusions: An improved method for testing the dynamic response of flow meters detects marked overshoot (underdamping) of TZ and MS responses not identified by the 26 ATS waveforms. This error could cause patient misclassification when using such meters with asthma guidelines.Keywords
This publication has 11 references indexed in Scilit:
- Cardiac troponin T in the severity assessment of patients with pulmonary embolism: cohort studyBMJ, 2003
- Recording flow in the first second of a maximal forced expiratory manoeuvre: influence of frequency contentEuropean Respiratory Journal, 2002
- Peak Expiratory Flow Profiles Delivered by Pump SystemsAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Method to produce American Thoracic Society flow-time waveforms using a mechanical pumpEuropean Respiratory Journal, 1997
- Peak expiratory flow and the resistance of the mini-wright peak flow meterEuropean Respiratory Journal, 1996
- Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan.Thorax, 1996
- Standardization of Spirometry, 1994 Update. American Thoracic Society.American Journal of Respiratory and Critical Care Medicine, 1995
- Portable peak flow meters: physical characteristics, influence of temperature, altitude, and humidityEuropean Respiratory Journal, 1994
- Accuracy of mini peak flow meters in indicating changes in lung function in children with asthmaBMJ, 1994
- The accuracy of portable peak flow meters.Thorax, 1992