Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressure
- 1 March 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Journal Of Hypertension
- Vol. 21 (3) , 571-576
- https://doi.org/10.1097/00004872-200303000-00023
Abstract
The Sphygmocor system purports to be able to assess ascending aortic blood pressure using a transfer function. It has been shown to be accurate when data obtained invasively are used, but has not been tested prospectively using data obtained non-invasively. To investigate the accuracy of this equipment when measurements are obtained non-invasively, as would normally be the case in the clinic setting. The study was observational. Ascending aortic pressure measurements were taken simultaneously with radial artery pressure wave recordings for estimation of ascending aortic blood pressure, in 28 patients undergoing diagnostic cardiac catheterization. The transfer function in the Sphygmocor system significantly underestimated invasively measured systolic blood pressure [mean -7.23 +/- 10.07 mmHg; 95% confidence interval (CI) -3.3 to -11.14 mmHg, P = 0.001] and significantly overestimated invasively measured diastolic blood pressure (mean 12.20 +/- 7.14 mmHg; 95% CI 9.43 to 14.97 mmHg, P <0.001). Oscillometrically measured brachial systolic blood pressure was not significantly different from that measured invasively in the ascending aorta (mean 3.36 +/- 10.47 mmHg; 95% CI -0.69 to 7.43 mmHg, P = 0.1), but oscillometric measurement of brachial diastolic blood pressure gave a significant overestimation of that measured invasively (mean 11.70 +/- 7.18 mmHg; 95% CI 8.91 to 14.49 mmHg, P <0.001). CONCLUSIONS The transfer function in the Sphygmocor system is no better at estimating ascending aortic blood pressure than are standard peripheral blood pressure measurements. It may be necessary to derive a new transfer system that is based on data that are acquired entirely non-invasively.Keywords
This publication has 19 references indexed in Scilit:
- Prospective Evaluation of a Method for Estimating Ascending Aortic Pressure From the Radial Artery Pressure WaveformHypertension, 2001
- Where is the evidence that radial artery tonometry can be used to accurately and noninvasively predict central aortic blood pressure in patients with diabetes?Diabetes Care, 2000
- Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometryIEEE Transactions on Biomedical Engineering, 1999
- Functional characterization of endothelin receptors in hypertensive resistance vesselsJournal Of Hypertension, 1999
- Non-invasive measurements of arterial structure and function: repeatability, interrelationships and trial sample sizeClinical Science, 1998
- Reproducibility of pulse wave velocity and augmentation index measured by pulse wave analysisJournal Of Hypertension, 1998
- Estimation of Central Aortic Pressure Waveform by Mathematical Transformation of Radial Tonometry PressureCirculation, 1997
- An analysis of the relationship between central aortic and peripheral upper limb pressure waves in manEuropean Heart Journal, 1993
- Does Radial Artery Pressure Accurately Reflect Aortic Pressure?Chest, 1992
- Nitroglycerin has more favourable effects on left ventricular afterload than apparent from measurement of pressure in a peripheral arteryEuropean Heart Journal, 1990