The incidence and significance of rub marks on the stem of removed total hip replacements
- 1 January 1983
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 17 (1) , 83-89
- https://doi.org/10.1002/jbm.820170108
Abstract
This article presents data on the incidence of rub marks (burnishing) on the stems of 180 removed total hip replacement femoral components most of which were McKees and Stanmores. The incidence varied with the reason for removal. It was low for assorted removals excluding fractured stem removals, higher for “loose stem” removals, and highest for fractured stem removals. For “loose stem” removals, it appeared to increase with implantation time. For fractured stem removals it appeared to increase with the time elapsed between fracture and removal. On not‐fractured removal, rub marks appeared to indicate that the component had been loose, but on fractured stem removals they appeared to indicate only that the component had been loose after fracture.This publication has 7 references indexed in Scilit:
- Analysis and prediction of total hip replacement performanceJournal of Materials Science, 1981
- Fracture and Corrosion in Stainless Steel Total Hip Replacement StemsPublished by ASTM International ,1979
- The Incidence of Fracture in Total Hip Replacement Femoral ComponentsEngineering in Medicine, 1978
- Welds in the Stems of Cast Cobalt Chrome Femoral Components for Use in Total Hip ReplacementsEngineering in Medicine, 1978
- A stress analysis of the femoral stem in total hip prosthesesJournal of Bone and Joint Surgery, 1976
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975
- Failed femoral stems in total hip prostheses. A report of six casesJournal of Bone and Joint Surgery, 1975