The duration times of left ventricular assist system (LVAS) support have increased because of prolonged wait times for transplant and the more frequent use of devices for destination therapy. The HeartMate LVAS, the only device approved for bridge to transplant and destination therapy, has limited durability, making replacement increasingly necessary. Since 1996, we have exchanged 19 left ventricular assist devices in 15 patients (11 men: mean age, 57.1 years; range, 33–77 years). Most of the devices (14) were replaced with the HeartMate vented electric/extended-lead vented electric pump; five devices were exchanged for a HeartMate II LVAS. Bearing failure was the most frequent reason for exchange (15 of 19 pumps); four of the 19 pumps also had active device-related infections at the time of exchange. There were no early deaths (30 days). Overall survival (Kaplan-Meier) was 85% at 1 year, 67% at 2 years, and 56% at 3 years. Three patients had transplants (mean, 518 days); six patients died during support (mean, 934 days), and six patients remain on LVAS support (mean, 1,219 days). One patient has been on device for over 6 years. Left ventricular assist devices exchange is becoming increasingly likely and can be associated with acceptably low-operative mortality rates and good intermediate-term survival.Keywords
This publication has 12 references indexed in Scilit:
- Outcomes of Left Ventricular Assist Device Implantation as Destination Therapy in the Post-REMATCH EraCirculation, 2007
- Management of Multiple Left Ventricular Assist Device Failures in a PatientThe Journal of Heart and Lung Transplantation, 2007
- Improved Mechanical Reliability of the HeartMate XVE Left Ventricular Assist SystemThe Annals of Thoracic Surgery, 2006
- Improved Durability of the HeartMate XVE Left Ventricular Assist Device Provides Safe Mechanical Support up to 1 Year But Is Associated With High Risk of Device Failure in the Second YearThe Journal of Heart and Lung Transplantation, 2006
- Assessment of Recurrent Heart Failure Associated with Left Ventricular Assist Device DysfunctionThe Journal of Heart and Lung Transplantation, 2005
- Left Ventricular Assist Device MalfunctionJournal of the American College of Cardiology, 2005
- HeartMate® VE LVAS design enhancements and its impact on device reliability1European Journal of Cardio-Thoracic Surgery, 2004
- Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantationThe Journal of Thoracic and Cardiovascular Surgery, 2001
- Long-Term Use of a Left Ventricular Assist Device for End-Stage Heart FailureNew England Journal of Medicine, 2001