Abstract
In standard teletherapy, a treatment plan is generated with the aid of a treatment planning system, but it is common to perform an independent monitor unit verification calculation (MUVC). In exact analogy, we propose and demonstrate that a simple and accurate MUVC in intensity modulated radiotherapy (IMRT) is possible. We introduce the concept of modified Clarkson integration (MCI). In MCI, we exploit the rotational symmetry of scattering to simplify the dose calculation. For dose calculation along a central axis (CAX), we first replace the incident IMRT fluence by an azimuthally averaged fluence. Second, the Clarkson integration is carried over annular sectors instead of over pie sectors. We wrote a computer code, implementing the MCI technique, in order to perform a MUVC for IMRT purposes. We applied the code to IMRT plans generated by CORVUS. The input to the code consists of CORVUS plan data (e.g., DMLC files, jaw settings, MU for each IMRT field, depth to isocenter for each IMRT field), and the output is dose contribution by individual IMRTs field to the isocenter. The code uses measured beam data for Sc, Sp, TPR, (D/MU)ref and includes effects from multileaf collimator transmission, and radiation field offset. On a 266 MHz desktop computer, the code takes less than 15 to calculate a dose. The doses calculated with the MCI algorithm agreed within +/-3% with the doses calculated by CORVUS, which uses a 1 cm x 1 cm pencil beam in dose calculation. In the present version of MCI, skin contour variations and inhomogeneities were neglected.Keywords
This publication has 33 references indexed in Scilit:
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- Testing of dynamic multileaf collimationMedical Physics, 1996
- Implementation of three dimensional conformal radiation therapy: Prospects, opportunities, and challengesInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Dose calculation for photon beams with intensity modulation generated by dynamic jaw or multileaf collimationsMedical Physics, 1994
- Characterization of a multileaf collimator systemInternational Journal of Radiation Oncology*Biology*Physics, 1993
- Three-dimensional conformal radiation therapy at the Memorial Sloan-Kettering Cancer CenterSeminars in Radiation Oncology, 1992
- Evaluation of multileaf collimator design for a photon beamInternational Journal of Radiation Oncology*Biology*Physics, 1992
- Methods of image reconstruction from projections applied to conformation radiotherapyPhysics in Medicine & Biology, 1990
- Description of the scatter component in photon-beam dataPhysics in Medicine & Biology, 1988
- A Note on Depth Doses in Fields of Irregular ShapeThe British Journal of Radiology, 1941