Improvement in Signal-to-noise Ratio and Reduction of Chemical Shift and Motion-induced Artifacts by Summation of Gradient and Spin Echo Data Acquisition
- 1 November 1989
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 13 (6) , 1041-1047
- https://doi.org/10.1097/00004728-198911000-00017
Abstract
Narrow bandwidth magnetic resonance (MR) imaging allows an increase of signal-to-noise ratio (SNR) but causes increased chemical shift and motion-induced artifacts. To obtain MR images with SNR approximately equal to that obtained with narrow bandwidth but with less chemical shift and motion-induced artifact, we introduced triple readout gradient reversal centered around the spin echo. As a result, signals from two gradient echoes and a single spin echo can be collected and summed. Phantom, knee, shoulder, and abdominal MR images were obtained using a 1.5 T GE Signa System at sampling rates ranging from 10 to 60 kHz. Since the bandwidth per pixel was tripled, chemical shift misregistration was reduced by the same factor. The summation image of two gradient echoes and one spin echo had an SNR comparable with that of a single spin echo acquired within the same total sampling interval. Data acquisition at a high sampling ratio also minimizes the dispersion of T2* weighting among three echoes. In addition, summation of the three resulting images decreases motion artifact by effective averaging.Keywords
This publication has 7 references indexed in Scilit:
- Motion artifact reduction in MR imaging of the abdomen: gradient moment nulling versus respiratory-sorted phase encoding.Radiology, 1988
- Multiple spin-echo MR imaging of the body: Image contrast and motion-induced artifactMagnetic Resonance Imaging, 1988
- Potential false-negative MR images of the thoracic spine in disk disease with switching of phase- and frequency-encoding gradients.Radiology, 1987
- Effect of the sampling rate on magnetic resonance imagingMagnetic Resonance in Medicine, 1987
- Motion Artifact Suppression Technique (MAST) for MR ImagingJournal of Computer Assisted Tomography, 1987
- Contrast in Rapid MR ImagingJournal of Computer Assisted Tomography, 1987
- Frequency Shift Artifacts in MR ImagingJournal of Computer Assisted Tomography, 1985