Tympanic membrane temperature as a measure of core temperature
Open Access
- 1 March 1999
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood
- Vol. 80 (3) , 262-266
- https://doi.org/10.1136/adc.80.3.262
Abstract
BACKGROUND Ear thermometers are becoming popular as a method for measuring deep body (core) temperature. AIM To determine the variability of a single user’s tympanic membrane (ear) temperature measurements. SUBJECTS Forty two, afebrile, healthy children, and 20 febrile children with acute burns. RESULTS In afebrile children measurements made in both ears (and within just a few minutes of each other) differed by as much as 0.6°C. Operator measurement error, sw of three consecutive measurements, in the same ear, was 0.13°C. In the group of febrile, burned children, core temperature was measured hourly at a number of sites (ear, rectum, axilla, bladder). A peak in core temperature occurred approximately 10–12 hours after the burn. Measurement error was calculated in 14 febrile, burned children with a peak temperature in excess of 38°C. For the left ear, measurement error was 0.19°C and for the right ear, 0.11°C. In the febrile children agreement between the ears was poor. The limits of agreement were 0.4°C to −0.8°C. It was not possible to predict the occasions when the temperature differences between the ears would be large or small. CONCLUSIONS The measurement error of one recording from the next is probably acceptable at about 0.1 to 0.2°C. To limit the variations in temperature of one ear to the other, measurements should be restricted to one of the ears whenever possible and the same ear used throughout the temperature monitoring period. Nurses and parents should take more than one temperature reading from the same ear whenever possible.Keywords
This publication has 13 references indexed in Scilit:
- A difference of 5°C between ear and rectal temperatures in a febrile patientThe American Journal of Emergency Medicine, 1997
- Statistics notes: Measurement errorBMJ, 1996
- Exercise-Induced Hyperthermia May Prevent Accurate Core Temperature Measurement by Tympanic Membrane ThermometerJournal of Nursing Measurement, 1996
- Acute changes in oxygen consumption and body temperature after burn injury.Archives of Disease in Childhood, 1994
- Evaluation of a tympanic thermometer in childrenThe Lancet, 1993
- Use of Tympanic Thermometers to Screen for Fever in Patients in a Pediatric Emergency DepartmentSouthern Medical Journal, 1993
- Relationships between skin temperature and perfusion in the arm and legClinical Physiology and Functional Imaging, 1991
- Assessment of an Aural Infrared Sensor for Body Temperature Measurement in ChildrenClinical Pediatrics, 1990
- Fever in burned childrenBurns, 1988
- DIAGNOSIS OF ACCIDENTAL HYPOTHERMIA OF THE ELDERLYThe Lancet, 1971