Liposome flow injection immunoassay: implications for sensitivity, dynamic range, and antibody regeneration
- 1 December 1990
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 62 (23) , 2587-2593
- https://doi.org/10.1021/ac00222a013
Abstract
We have developed a liposome-based flow injection immunoassay (FIIA) system for quantitation of a clinical analyte, theophylline. With very minor changes in assay format, this procedure can also be used for the quantitation of anti-theophylline. Automated sequential analyses were performed at room temperature with picomole sensitivity and a day-to-day coefficient of variation of less than 5% for aqueous solutions. The system components include liposomes that contain fluorophores in their aqueous centers and an immobilized-antibody reactor column. The immunoractor was regenerated hundreds of times over 3 months of continuous use with no measurable loss of antibody activity. The two assay formats studied produced distinct dynamic ranges for their respective analytes. The special advantages of using flow injection analysis for immunoassays and of using liposomes in FIIA are discussed.This publication has 7 references indexed in Scilit:
- Generic liposome reagent for immunoassaysAnalytical Biochemistry, 1989
- Behavior of liposomes in flow injection systemsAnalytical Chemistry, 1988
- Extending the detection limit of solid-phase electrochemical enzyme immunoassay to the attomole levelAnalytical Biochemistry, 1988
- Liposome immunosensor for theophyllineBiochemical and Biophysical Research Communications, 1980
- The adsorptive characteristics of proteins for polystyrene and their significance in solid-phase immunoassaysAnalytical Biochemistry, 1980
- Statistical estimation of the minimal detectable concentration (“sensitivity”) for radioligand assaysAnalytical Biochemistry, 1978
- Human Red Cell Glycolytic IntermediatesJournal of Biological Chemistry, 1959