Does Hematocrit Affect In Vitro Hemolysis Test Results? Preliminary Study with Baylor/NASA Prototype Axial Flow Pump
- 1 September 1994
- journal article
- Published by Wiley in Artificial Organs
- Vol. 18 (9) , 650-656
- https://doi.org/10.1111/j.1525-1594.1994.tb03394.x
Abstract
The effect of hematocrit (Ht) on in vitro hemolysis test results (i. e., index of hemolysis) was evaluated using a Baylor/NASA prototype axial flow pump. Red blood cell suspensions of six different Ht (5, 10, 15, 20, 30, 40%; n = 30) were prepared and used for this evaluation. The pump was operated for 60 min under 5 L/min flow conditions, and blood samples were taken every 10 min to measure plasma free hemoglobin levels. The normalized index of hemolysis (NIH) was calculated using the regression line slope between time and plasma free hemoglobin level, and relationships between NIH and Ht or hemoglobin (Hb) were checked. NIH and Ht had a statistically significant (p < 0.0001) correlation with a coefficient of fit of 0.976, and NIH and Hb had a statistically significant (p < 0.0001) correlation with a coefficient of fit of 0.976. To reduce the effect of Ht, NIH/Ht was pro posed and compared with a modified index of hemolysis (MIH), which was normalized by the Hb level of blood. NIH/Ht and MIH had a poor correlation with Ht (coefficient of fit, 0.608) and Hb (coefficient of fit, 0.577), respectively. When blood that has a wide range of Ht or Hb values is used for in vitro hemolysis tests, NIH/Ht is suggested for use as an index of hemolysis to evaluate the hemolysis characteristics of rotary blood pumps because MIH has no dimension and it requires Hb values. In contrast, NIH/Ht has a dimension of g/100 L, which is quite understandable, and it does not require measurement of Hb levels of blood; it is therefore cost–effective.Keywords
This publication has 8 references indexed in Scilit:
- The Need for Standardizing the Index of HemolysisArtificial Organs, 1994
- Development of the Baylor/NASA Axial Flow Ventricular Assist Device: In Vitro Performance and Systematic Hemolysis Test ResultsArtificial Organs, 1994
- In Vitro Performance of the Baylor/NASA Axial Flow PumpArtificial Organs, 1993
- In Vitro Hematological Testing of Rotary Blood Pumps: Remarks on Standardization and Data InterpretationArtificial Organs, 1993
- A Compact Centrifugal Pump for Cardiopulmonary BypassArtificial Organs, 1992
- Red Blood Cell Damage by Shear StressBiophysical Journal, 1972
- Contribution to the in vitro testing of pumps for extracorporeal circulationThe Journal of Thoracic and Cardiovascular Surgery, 1967
- Studies of Plasma Hemoglobin Formation in three Pumping Systems Used in Extracorporeal CirculationAnnals of Surgery, 1962