Use of the Reflotron system for cholesterol assay in capillary blood, venous blood, and serum--evaluation of accuracy and lot-to-lot reagent comparability.
Open Access
- 1 October 1988
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 34 (10) , 2117-2119
- https://doi.org/10.1093/clinchem/34.10.2117
Abstract
Since its recalibration in 1987, the Reflotron "dry-chemistry" cholesterol method can produce accurate results. Here we report how cholesterol concentrations in venous blood, plasma, and serum determined with Reflotron compare with values determined for fingerstick samples. When no EDTA is used in the latter, accurate results may meet the WHO/CDC criterion of a maximum bias less than or equal to 5%. Reflotron users should avoid use of EDTA. Lot-to-lot variation among the solid-phase reagent carriers was evaluated in four lots of these strips. The largest difference, 0.4 mmol/L, was seen in a commercial quality-control serum with a cholesterol concentration of approximately 7 mmol/L. The maximum differences in results for several serum pools were not always found between the same reagent lots, although the highest results in five of six materials occurred with a single lot of strips. As in all clinical chemical testing, regular quality control is necessary.This publication has 8 references indexed in Scilit:
- Revised calibration of the Reflotron cholesterol assay evaluated.Clinical Chemistry, 1988
- A MULTICENTER STUDY OF THE NEW REFLOTRON SYSTEM FOR THE MEASUREMENT OF UREA, GLUCOSE, TRIACYLGLYCEROLS, CHOLESTEROL, GAMMA-GLUTAMYLTRANSFERASE AND HEMOGLOBIN1988
- Bias between enzymatic methods and the reference method for cholesterol.Clinical Chemistry, 1988
- Comparison of Two Solid Phase Chemistry Systems: Reflotron and Ektachem DT 60cclm, 1987
- Interlaboratory testing of the transferability of a candidate reference method for total cholesterol in serum.Clinical Chemistry, 1986
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part Icclm, 1983
- Matrix effects in clinical analysis: commutability of control materials between the Ektachem, Beckman and SMA glucose and urea methodsClinica Chimica Acta; International Journal of Clinical Chemistry, 1981
- Simultaneously obtained skin-puncture serum, skin-puncture plasma, and venous serum compared, and effects of warming the skin before puncture.Clinical Chemistry, 1977