Absolute Calibration of Light Sources in the Vacuum Ultraviolet by Means of the Synchrotron Radiation of DESY
- 1 February 1969
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 8 (2) , 255-259
- https://doi.org/10.1364/ao.8.000255
Abstract
With the synchrotron radiation of DESY (Deutsches Elektronen Synchrotron) as a primary standard light source, secondary standard light sources have been calibrated in the spectral range 1500–2700 Å. By this method, the accuracy of the intensity of the calibrated lamp is about 4%.Keywords
This publication has 4 references indexed in Scilit:
- The Synchrotron Radiation of the 6-GeV DESY Machine as a Fundamental Radiometric StandardApplied Optics, 1967
- Die räumliche Strahlungsverteilung des UV-StandardsAnnalen der Physik, 1952
- On the Classical Radiation of Accelerated ElectronsPhysical Review B, 1949