Correlation Between Airflow Patterns and Performance of a Laboratory Fume Hood
- 1 December 2006
- journal article
- research article
- Published by Taylor & Francis in Journal of Occupational and Environmental Hygiene
- Vol. 3 (12) , 694-706
- https://doi.org/10.1080/15459620601015695
Abstract
To understand the physical mechanisms of the contaminant dispersion and containment leakage during the ventilation process through a laboratory fume hood, the complicated three-dimensional flow patterns and the real-time tracer gas (SF6) leakage were studied via the laser-assisted flow visualization method and the standard/special gas sampling technique, respectively. Through flow visualization, the large-scale vortex structures and boundary layer separations were found around the side poles and doorsill of the hood. In the near-wake region of the manikin, large recirculation zones and wavy flow structures were also identified. When tracer gas concentration measurements were conducted point-by-point across the sash opening, the areas near the doorsill, the lower parts of the side poles, and the sides of the manikin showed significant contaminant leaks. These areas with high contaminant leaks exactly corresponded to where the flow recirculated or separated. However, when the ANSI/ASHRAE 110–1995 protocol was used to measure the concentration of SF6 at the breathing zone of the manikin, no appreciable leakage was detected. It is suggested that a method based on the aerodynamic features and multipoint leakage detections would reflect a more realistic evaluation of overall performance of laboratory fume hood than a single-point sampling method at the manikin's breathing zone.Keywords
This publication has 15 references indexed in Scilit:
- Required response time for variable air volume fume hood controllersAnnals of Occupational Hygiene, 2000
- Fume Hood Performance: Face Velocity Variability Inconsistent Air Volume SystemsApplied Occupational and Environmental Hygiene, 1998
- Effect of Baffles and a Louvered Bypass on the Airflow and the Convective Patterns of Contaminant Inside a Fume HoodAihaj Journal, 1998
- Reducing Employee Exposure Potential Using the ANSI/ASHRAE 110 Method of Testing Performance of Laboratory Fume Hoods as a Diagnostic ToolAihaj Journal, 1998
- EFFECT OF THE LOCATION OF THE EXHAUST DUCT, AN EXTERIOR OBSTRUCTION AND HANDLE ON THE AIR FLOW INSIDE AND AROUND A FUME CUPBOARDAnnals of Work Exposures and Health, 1996
- CONTAINMENT TESTING OF FUME CUPBOARDS—II. TEST ROOM MEASUREMENTSAnnals of Work Exposures and Health, 1992
- Numerical simulation and optimization of the aerodynamics flow field inside fume cupboardsJournal of Wind Engineering and Industrial Aerodynamics, 1992
- CONTAINMENT TESTING OF FUME CUPBOARDS—I.METHODSAnnals of Work Exposures and Health, 1992
- Experimental and numerical investigations of the performance of fume cupboardsBuilding and Environment, 1991
- A New Method for Quantitative, In-Use Testing of Laboratory Fume HoodsAihaj Journal, 1989