Effect of Air Space on Depth Dose in Electron Beam Therapy
- 1 January 1978
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 126 (1) , 249-251
- https://doi.org/10.1148/126.1.249
Abstract
In electron beam turmor therapy, alterations in dosimetry occur as a result of air space between the end-of-treatment cone and the skin surface. A large air gap may be introduced to obtain a field size larger than that available at the cone end. Needed dosimetry corrections related to these air space problems are discussed, along with a proposed method of measuring effective source-to-cone end distance. Data presented show the modifications of a dosimetric field which occur with an increase in the air space below the treatment cone.Keywords
This publication has 1 reference indexed in Scilit:
- Field shaping in electron beam therapyThe British Journal of Radiology, 1976