Skin thickness measurement by pulsed ultrasound: its reproducibility, validation and variability
- 1 June 1982
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Dermatology
- Vol. 106 (6) , 657-667
- https://doi.org/10.1111/j.1365-2133.1982.tb14702.x
Abstract
The reproducibility of the pulsed ultrasound technique for the determination of skin thickness was investigated, using two independent observers. No systematic difference was found and a high correlation was obtained. Studies were also undertaken to validate the pulsed ultrasound technique as a measure of true skin thickness. Skin thickness determined in vitro was found to be greater than when in vivo determinations were made by either the pulsed ultrasound or a xeroradiographic technique, probably due to the release of in vivo tension within the dermis after excision. Skin thickness was found to increase linearly with age up to the age of 20 years and to decrease linearly with age subsequently. Differences in skin thickness between the sexes and in different sites of the body were demonstrated.This publication has 7 references indexed in Scilit:
- Comparison of Xeroradiographic and Ultrasound Detection of Corticosteroid Induced Dermal ThinningJournal of Investigative Dermatology, 1981
- Xeroradiographic and ultrasound techniques in the assessment of skin disorderPublished by Springer Nature ,1981
- Determining Skin Thickness with Pulsed Ultra SoundJournal of Investigative Dermatology, 1979
- Roentgenographic Measurement of Skin Thickness in Normal IndividualsRadiology, 1975
- The measurement of corticosteroid induced dermal atrophy by a radiological methodArchives of Dermatological Research, 1975
- A MODIFIED RADIOGRAPHIC METHOD FOR MEASURING SKIN THICKNESS*British Journal of Dermatology, 1969
- Roentgenographic Visualization and Measurement of SkinRadiology, 1964