Polyethylene wear of the PCA unicompartmental knee
- 1 January 1992
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 63 (3) , 260-262
- https://doi.org/10.3109/17453679209154778
Abstract
In Lund, 120 arthrosis knees had the PCA unicompartmental endoprosthesis during 1983-1987 and were followed prospectively for 5 (4-8) years. 17 knees were excluded from the final clinical follow-up because of the development of rheumatoid arthritis (2), severe neurologic disease (3), or death (12). Subjectively, 68 knees were much improved, 26 improved and 9 had failed. First-steps problems were only present in the uncemented group of 49 knees. Of the 9 failures (7 cemented and 2 uncemented) 6 were revised. The main reason for revision was loosening of the femoral component (2), tibial component (2), or both components (1), and polyethylene wear (1). However, all revised tibial components showed polyethylene wear, in 4 quite pronounced. Weight-bearing radiographs (61 knees) revealed major polyethylene wear in an additional 14 knees; not only in thin tibial components. There may be an increasing clinical problem due to polyethylene wear. With the presented findings the PCA unicompartmental knee endoprosthesis cannot be recommended.Keywords
This publication has 10 references indexed in Scilit:
- Multicenter study of unicompartmental knee revisionActa Orthopaedica, 1992
- Investigation of Early Surface Delamination Observed in Retrieved Heat-Pressed Tibial InsertsClinical Orthopaedics and Related Research, 1991
- Cementless PCA unicompartmental joint arthroplasty for osteoarthritis of the knee: A prospective study of 51 casesThe Journal of Arthroplasty, 1990
- Polyethylene failure in two total knees: Wear of thin, metal-backed PCA tibial componentsActa Orthopaedica, 1990
- 2 bis 4 Jahresresultate der Zementfreien Anwendung der PCA Unicondylären KnieendoprotheseZeitschrift für Orthopädie und ihre Grenzgebiete, 1990
- The PCA unicompartmental knee: A 1–4-year comparison of fixation with or without cementActa Orthopaedica, 1988
- The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement.Journal of Bone and Joint Surgery, 1986
- The Effect of Conformity and Plastic Thickness on Contact Stresses in Metal-Backed Plastic ImplantsJournal of Biomechanical Engineering, 1985
- Deformation and Loosening of the Tibial Component in Knee Arthroplasty with Unicompartmental EndoprosthesesActa Orthopaedica, 1981