Deactivation of fused silica capillary columns with phenylhydrosiloxanes
- 1 August 1985
- journal article
- research article
- Published by Wiley in Journal of High Resolution Chromatography
- Vol. 8 (8) , 378-384
- https://doi.org/10.1002/jhrc.1240080803
Abstract
In this work, an investigation of new organosilicon hydride reagents with phenyl functional groups for deactivation and surface modification of fused silica capillary columns is described. Different reagents were tested for their ability to deactivate the fused silica surface, and actual critical surface tension measurements were made using the capillary rise method. The deactivation procedure required lower optimum temperatures than conventional methods. Deactivated capillaries and coated capillary columns were prepared and tested for reproducibility, efficiency, and surface inertness towards basic and acidic compounds at the low nanogram level.Keywords
This publication has 15 references indexed in Scilit:
- A CP‐MAS NMR study of some deactivation methods in capillary gas chromatographyJournal of High Resolution Chromatography, 1984
- Deactivation of fused silica capillary columns with polymethylhydrosiloxanesJournal of High Resolution Chromatography, 1984
- Contemporary capillary columns for gas chromatographyJournal of High Resolution Chromatography, 1984
- Polarizable biphenyl polysiloxane stationary phase for capillary column gas chromatographyJournal of High Resolution Chromatography, 1984
- Determination of surface hydroxyl concentration on glass and fused silica capillary columnsChromatographia, 1982
- Characterization of glass, quartz, and fused silica capillary column surfaces from contact-angle measurementsChromatographia, 1981
- Deactivation of glass capillaries by persilylation. Part 3: Extending the wettability by bonding phenyl groups to the glass surfaceJournal of High Resolution Chromatography, 1980
- Deactivation of glass capillaries by persilylation. Part 2: Practical recommendationsJournal of High Resolution Chromatography, 1979
- Alkylpolysiloxane glass capillary columns combining high temperature stability of the stationary liquid and deactivation of the surfaceChromatographia, 1979
- Zur Desaktivierung von Glaskapillaren mittels SilanisierungChromatographia, 1977