Stabilization of High-Resistance Seals in Patch-Clamp Recordings by Laminar Flow
- 22 October 2003
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 75 (23) , 6718-6722
- https://doi.org/10.1021/ac0346611
Abstract
The formation of a high-resistance electrical seal between a cell membrane and a glass micropipet tip is essential in patch-clamp experiments. We have studied the electrical properties and the mechanical stability of the seal using a microfluidic chip generating laminar flow in open volumes. We show that, by using fluid flow (1−10 mm/s) acting along the symmetry axis of the cell−pipet, seals of a higher mechanical stability with increased resistances can be achieved, allowing up to 100% longer recording times and over 40% decreased noise levels (Irms). These improved properties are beneficial for high-sensitivity patch-clamp recordings, in particular, in longtime studies of ion channel receptor systems that are relevant in biosensor applications of the technique. Furthermore, these observations support the combination of patch-clamp with microfluidic devices, for example, for rapid solution exchange around a single cell sensor for high-throughput electrophysiology and for highly resolved kinetic studies.Keywords
This publication has 12 references indexed in Scilit:
- A Cell-Based Bar Code Reader for High-Throughput Screening of Ion Channel−Ligand InteractionsAnalytical Chemistry, 2002
- Novel animal-health drug targets from ligand-gated chloride channelsNature Reviews Drug Discovery, 2002
- Microsecond Time-Scale Discrimination Among Polycytidylic Acid, Polyadenylic Acid, and Polyuridylic Acid as Homopolymers or as Segments Within Single RNA MoleculesPublished by Elsevier ,1999
- The ion channel switch biosensorJournal of Molecular Recognition, 1999
- [10] Ion channels as tools to monitor lipid bilayer-membrane protein interactions: Gramicidin channels as molecular force transducersPublished by Elsevier ,1999
- Supported Membranes: Scientific and Practical ApplicationsScience, 1996
- Physical Actions in Biological AdhesionPublished by Elsevier ,1995
- Pharmacological properties of the homomeric α7 receptorNeuroscience Letters, 1992
- Open channel noise. I. Noise in acetylcholine receptor currents suggests conformational fluctuationsBiophysical Journal, 1985
- A reinterpretation of mammalian sodium channel gating based on single channel recordingNature, 1983