A SIMPLE GAMMA-RAY INSENSITIVE FAST-NEUTRON COUNTER
- 1 July 1959
- journal article
- Published by Canadian Science Publishing in Canadian Journal of Physics
- Vol. 37 (7) , 858-865
- https://doi.org/10.1139/p59-093
Abstract
A gamma-ray insensitive fast-neutron counter is described which consists of a number of thin rectangular sheets of lucite coated with zinc sulphide and sandwiched together to form a rectangular block which is mounted on a photo-multiplier. Pulse height spectra and absolute efficiency curves are presented for neutrons from 300 kev to 15 Mev.Keywords
This publication has 12 references indexed in Scilit:
- Neutron Spectra from and Using Pulse Shape Discrimination in an Organic ScintillatorPhysical Review Letters, 1959
- Monoenergetic Neutron Techniques in the 10- to 30-Mev RangeReviews of Modern Physics, 1956
- ReactionPhysical Review B, 1955
- Detection of 2.5-Mev Neutrons with a Single-Crystal Scintillation CounterReview of Scientific Instruments, 1954
- A Fast Neutron ScintillatorReview of Scientific Instruments, 1954
- A Fast Neutron DetectorReview of Scientific Instruments, 1952
- Yield and Angular Distribution of the D-D NeutronsPhysical Review B, 1949
- The Use of Anthracene as a Scintillation CounterPhysical Review B, 1948
- A Neutron Detector Having Uniform Sensitivity from 10 Kev to 3 MevPhysical Review B, 1947
- Hydrogen Recoil Proportional Counter for Neutron DetectionReview of Scientific Instruments, 1947