Quantitative analysis of image truncation in focal-beam CT
- 1 October 1993
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 38 (10) , 1443-1457
- https://doi.org/10.1088/0031-9155/38/10/006
Abstract
The authors measured truncation artifacts using a phantom model of the human thorax. Reconstructions were performed with both iterative cone-beam transmission maximum likelihood (TRML) and cone-beam filtered backprojection (CB-FBP). When CB-FBP was used, image truncation created 'streak' artifacts of increased density, distorted the elliptical phantom body contour into a circle with reduced radius, and increased the apparent density of the lungs. SPECT attenuation compensation using this truncated CB-FBP map produced line-source activities that were only 65% of the correct 'in-air' values. Thus, CB-FBP attenuation maps will generally be unacceptable for clinical SPECT attenuation compensation. The truncation-streak artifacts were eliminated with TRML, but in some regions the body contour was not sharply defined, and the reconstructed density was low in the lungs and nonzero outside the body contour. An elliptical support prior eliminated the 'in-air' density and improved the accuracy in the lungs, as determined by narrow-beam transmission measurements with a germanium detector. For iterations above twenty-four, TRML images were noisier than CB-FBP images with a Hann filter, and at least thirty iterations were required for accurate reconstruction. SPECT line-source attenuation compensation with a truncated TRML map was accurate to within 20%, depending upon the slice number and the source location.Keywords
This publication has 18 references indexed in Scilit:
- Attenuation and detector response compensations used with Gibbs prior distributions for maximum a posteriori SPECT reconstructionIEEE Transactions on Nuclear Science, 1992
- Truncation artifact suppression in cone-beam radionuclide transmission CT using maximum likelihood techniques: evaluation with human subjectsPhysics in Medicine & Biology, 1992
- Attenuation maps for SPECT determined using cone beam transmission computed tomographyIEEE Transactions on Nuclear Science, 1990
- Feasible images and practical stopping rules for iterative algorithms in emission tomographyIEEE Transactions on Medical Imaging, 1989
- Weighted backprojection implemented with a non-uniform attenuation map for improved SPECT quantitationIEEE Transactions on Nuclear Science, 1988
- Tc‐99m attenuation coefficients in water‐filled phantoms determined with gamma camerasMedical Physics, 1984
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984
- Evaluation of a Preprocessing Algorithm for Truncated CT ProjectionsJournal of Computer Assisted Tomography, 1981
- Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation camerasPhysics in Medicine & Biology, 1979
- Attenuation Correction and Incomplete Projection in Single Photon Emission Computed TomographyIEEE Transactions on Nuclear Science, 1979