A low temperature calibration equation for carbon resistance thermometers in high magnetic fields
- 1 January 1974
- journal article
- research article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 7 (1) , 36-38
- https://doi.org/10.1088/0022-3735/7/1/010
Abstract
An interpolation equation for a 470 Omega , 0.5 W Speer resistor has been found, which with 20 calibration points enables an interpolation accuracy of better than 3 mK to be obtained in the range 0.3-4 K and in magnetic fields of up to 6.5 T.Keywords
This publication has 13 references indexed in Scilit:
- Negative Magnetoresistance in CarbonJournal of the Physics Society Japan, 1973
- Thermometer equations for low-temperature glass-ceramic capacitance thermometersCryogenics, 1972
- A Low Temperature Glass-Ceramic Capacitance ThermometerReview of Scientific Instruments, 1971
- Studies of a Glass-Ceramic Capacitance Thermometer in an Intense Magnetic Field at Low TemperaturesReview of Scientific Instruments, 1971
- Studies of a Glass-Ceramic Capacitance Thermometer between 0.025 and 2.4 KReview of Scientific Instruments, 1971
- Low Temperature Thermometry in High Magnetic Fields. II. Germanium and Platinum ResistorsReview of Scientific Instruments, 1971
- Low Temperature Thermometry in High Magnetic Fields. I. Carbon ResistorsReview of Scientific Instruments, 1969
- Specific Heats below 3°K of Pure Copper, Silver, and Gold, and of Extremely Dilute Gold-Transition-Metal AlloysPhysical Review B, 1968
- A simple helium-3 cryostat for specific heat measurements between 0.4 and 4.2 KJournal of Scientific Instruments, 1965
- Low-temperature specific heat of dilute Cu-Mn alloysJournal of Physics and Chemistry of Solids, 1960