Abstract
Objective. To analyse the occurrence of hardware-related complications in patients with deep brain stimulation (DBS), over a long period of time. Method. All patients operated on with DBS at our institution between 1993 and 2002 were followed with respect to adverse events related to the implanted hardware. Results. One hundred and nineteen consecutive patients underwent 139 procedures with implantation of 161 electrodes. The minimum follow-up was 12 months. The follow-up time was 540 electrode-years. The rate of hardware-related complications per electrode-year was 4.3%. In total, 17 patients (15%) had 23 hardware-related complications. These included 8 electrode breakages, 4 electrode migrations, 2 stimulator migrations, 3 erosions, 2 erosions and infections, 2 infections and 2 cases of stimulator malfunction. The majority of these complications occurred during the first four years in our experience. Conclusions. DBS is a life-long therapy that requires a life-long follow-up. Increased experience and adaptation of surgical technique are the main determinants for avoidance of hardware-related complications.Keywords
This publication has 10 references indexed in Scilit:
- Technical Complication in Deep Brain StimulationCentral European Neurosurgery, 2002
- Hardware-Related problems of deep brain stimulationMovement Disorders, 2002
- Hardware-Related Complications after Placement of Thalamic Deep Brain Stimulator SystemsStereotactic and Functional Neurosurgery, 2002
- Revision of deep brain stimulator for tremorJournal of Neurosurgery, 2001
- Hardware failure in parkinsonian patients with chronic subthalamic nucleus stimulation is a medical emergencyMovement Disorders, 2001
- Complications of Deep Brain Stimulation SurgeryStereotactic and Functional Neurosurgery, 2001
- A Comparison of Continuous Thalamic Stimulation and Thalamotomy for Suppression of Severe TremorNew England Journal of Medicine, 2000
- Deep brain stimulation is preferable to thalamotomy for tremor suppressionSurgical Neurology, 1998
- Comparison between Stereotactic CT and MRI Coordinates of Pallidal and Thalamic Targets using the Laitinen Noninvasive StereoadapterStereotactic and Functional Neurosurgery, 1998
- Deep Brain Stimulation and Thalamotomy for Tremor ComparedPublished by Springer Nature ,1997