Low-dose computed tomographic imaging in orbital trauma

Abstract
We review the findings in 75 computed tomographic (CT) examinations of 66 patients with orbital trauma who were imaged using a low-radiation-dose CT technique. Imaging was performed using a dynamic scan mode and exposure factors of 120 kVp and 80 mAs resulting in a skin dose of 11 mGy with an effective doseequivalent of 0.22 mSv. Image quality was diagnostic in all cases and excellent in 73 examinations. Soft-tissue abnormalities within the orbit including muscle adhesions were well demonstrated both on primary axial and reconstructed multiplanar images. The benefits of multiplanar reconstructions are stressed and the contribution of soft-tissue injuries to symptomatic diplopia is examined.