The optimization of signal detectability in digital fluoroscopy
- 1 September 1982
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 9 (5) , 688-694
- https://doi.org/10.1118/1.595130
Abstract
An optimization analysis of digital fluoroscopy was carried out using (signal/noise)2/dose as a figure of merit. Three signal detection tasks have been considered: small I-containing blood vessels; regions of fat; and microcalcifications. The effects of X-ray beam filtration and kilovoltage on the figure of merit were parameterized and the associated X-ray tube power loadings were defined. Results are presented in a fashion that facilitates the choice of parameters (amount of Al filtration, kilovoltage, and X-ray tube power loading) such that an acceptable balance can be achieved between signal detection sensitivity, patient dose and X-ray tube power loading.This publication has 1 reference indexed in Scilit:
- Lesion detection and signal–to–noise ratio in CT imagesMedical Physics, 1981