An improved analytical detector response function model for multilayer small-diameter PET scanners
- 1 April 2003
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (8) , 979-994
- https://doi.org/10.1088/0031-9155/48/8/302
Abstract
The optimization of spatial resolution is a critical consideration in the design of small-diameter positron emission tomography (PET) scanners for animal imaging, and is often addressed with Monte Carlo simulations. As a faster and simpler solution, we have developed a new analytical model of the PET detector response function, and implemented the model for a small single-slice, multilayer PET scanner. The accuracy of the model has been assessed by comparison with both Monte Carlo simulations and experimental measurements published in the literature. Results from the analytical model agreed well with the Monte Carlo method, being noise free and two to three orders of magnitude faster. The only major discrepancy was a slight underestimation of the width of the point spread function by the analytical method as inter-crystal scatter is neglected. We observed good agreement between the predictions of the model and experimental measurements. For two large-diameter scanners additional discrepancies were seen due to photon acollinearity, which is not considered in the model. We have shown that the simple and fast analytical detector response function model can provide accurate estimates of spatial resolution for small-diameter PET scanners, and could be a useful tool for several applications, complementing or cross-validating other simulation methods.Keywords
This publication has 24 references indexed in Scilit:
- A depth of interaction detector for PET with GSO crystals doped with different amounts of CeIEEE Transactions on Nuclear Science, 2002
- High spatial resolution small animal YAP-PETNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1998
- Use of scanner characteristics in iterative image reconstruction for high-resolution positron emission tomography studies of small animalsEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997
- PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat heartsNMR in Biomedicine, 1997
- Design and engineering aspects of a high resolution positron tomograph for small animal imagingIEEE Transactions on Nuclear Science, 1994
- Effect of detector weighting functions on the point spread function of high-resolution PET tomographs: a simulation studyIEEE Transactions on Medical Imaging, 1992
- Study of the resolution performance of an array of discrete detectors with independent readouts for positron emission tomographyIEEE Transactions on Medical Imaging, 1991
- Initial Results from the Donner 600 Crystal Positron TomographIEEE Transactions on Nuclear Science, 1987
- Analysis of Positron-Emission Tomography Scintillation-Detectors with Wedge Faces and Inter-Crystal SeptaIEEE Transactions on Nuclear Science, 1985
- PET Detector Ring Aperture Function Calculations Using Monte Carlo TechniquesIEEE Transactions on Nuclear Science, 1983