Nanoflow Solvent Gradient Delivery from a Microfabricated Device for Protein Identifications by Electrospray Ionization Mass Spectrometry
- 30 July 1998
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 70 (18) , 3721-3727
- https://doi.org/10.1021/ac980502j
Abstract
Microfabrication technology offers the opportunity to construct microfluidic modules which are designed to perform specific, dedicated functions. Here we report the construction of a microfabricated device for the generation and delivery by electroosmotic pumping of solvent gradients at nanoliter per minute flow rates. The device consists of three solvent reservoirs and channels which were etched in glass. Solvent gradients and solvent flows were generated by computer controlled differential electroosmotic pumping of aqueous and organic phase, respectively, from the solvent reservoirs. The device was integrated into an analytical system consisting of the solvent gradient delivery module, a reverse phase micro-column and an electrospray ionization ion trap mass spectrometer (MS). The system was used for the analysis at high sensitivity of peptides and peptide mixtures generated by proteolytic digestion of proteins. We have measured an absolute limit of detection as low as 1 fmol and a concentration limit of detection at the 100 amol/μL level. The system was also successfully used for the identification of proteins separated by 1D and 2D gel electrophoresis. This was achieved by gradient frontal analysis of the peptide mixture generated by proteolysis of the respective proteins, and the automated generation and interpretation of collision-induced dissociation spectra.Keywords
This publication has 25 references indexed in Scilit:
- Generating Electrospray from Microchip Devices Using Electroosmotic PumpingAnalytical Chemistry, 1997
- High sensitivity identification of proteins by electrospray ionization tandem mass spectrometry: Initial com‐ parison between an ion trap mass spectrometer and a triple quadrupole mass spectrometerElectrophoresis, 1997
- Protein Identification by Capillary Zone Electrophoresis/Microelectrospray Ionization-Tandem Mass Spectrometry at the Subfemtomole LevelAnalytical Chemistry, 1996
- Electrospray Ionization Mass Spectrometric Investigation of Signal Transduction PathwaysPublished by American Chemical Society (ACS) ,1996
- Instrumental Requirements for Nanoscale Liquid ChromatographyAnalytical Chemistry, 1996
- Femtomole sequencing of proteins from polyacrylamide gels by nano-electrospray mass spectrometryNature, 1996
- Tutorial review. Future prospects for the analysis of complex biological systems using micro-column liquid chromatography–electrospray tandem mass spectrometryThe Analyst, 1996
- A Microscale Electrospray Interface for Online, Capillary Liquid Chromatography/Tandem Mass Spectrometry of Complex Peptide MixturesAnalytical Chemistry, 1995
- Whole-Genome Random Sequencing and Assembly of Haemophilus influenzae RdScience, 1995
- High‐sensitivity determination of tyrosine‐phosphorylated peptides by on‐line enzyme reactor and electrospray ionization mass spectrometryProtein Science, 1995