Clinical evaluation of 2D versus 3D whole-body PET image quality using a dedicated BGO PET scanner
- 22 April 2005
- journal article
- research article
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 32 (9) , 1050-1056
- https://doi.org/10.1007/s00259-005-1809-9
Abstract
Three-dimensional positron emission tomography (3D PET) results in higher system sensitivity, with an associated increase in the detection of scatter and random coincidences. The objective of this work was to compare, from a clinical perspective, 3D and two-dimensional (2D) acquisitions in terms of whole-body (WB) PET image quality with a dedicated BGO PET system. 2D and 3D WB emission acquisitions were carried out in 70 patients. Variable acquisition parameters in terms of time of emission acquisition per axial field of view (aFOV) and slice overlap between sequential aFOVs were used during the 3D acquisitions. 3D and 2D images were reconstructed using FORE+WLS and OSEM respectively. Scatter correction was performed by convolution subtraction and a model-based scatter correction in 2D and 3D respectively. All WB images were attenuation corrected using segmented transmission scans. Images were blindly assessed by three observers for the presence of artefacts, confidence in lesion detection and overall image quality using a scoring system. Statistically significant differences between 2D and 3D image quality were only obtained for 3D emission acquisitions of 3 min. No statistically significant differences were observed for image artefacts or lesion detectability scores. Image quality correlated significantly with patient weight for both modes of operation. Finally, no differences were seen in image artefact scores for the different axial slice overlaps considered, suggesting the use of five slice overlaps in 3D WB acquisitions. 3D WB imaging using a dedicated BGO-based PET scanner offers similar image quality to that obtained in 2D considering similar overall times of acquisitions.Keywords
This publication has 16 references indexed in Scilit:
- Clinical validation of fully 3-D versus 2.5-D RAMLA reconstruction on the Philips-ADAC CPET PET scannerNuclear Medicine Communications, 2004
- Technology related parameters affecting quantification in positron emission tomography imagingNuclear Medicine Communications, 2004
- Performance evaluation of the new whole-body PET/CT scanner: Discovery STEuropean Journal of Nuclear Medicine and Molecular Imaging, 2004
- Impact of patient weight and emission scan duration on PET/CT image quality and lesion detectability.2004
- A lesion detection observer study comparing 2-dimensional versus fully 3-dimensional whole-body PET imaging protocols.2004
- Optimal dose of 18 F-FDG required for whole-body PET using an LSO PET cameraEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- Parameterization of a model-based 3-D PET scatter correctionIEEE Transactions on Nuclear Science, 2002
- Optimization of injected dose based on noise equivalent count rates for 2- and 3-dimensional whole-body PET.2002
- Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the ProjectionsJournal of Computer Assisted Tomography, 1983