Aortic Valve Replacement with the Carpentier-Edwards Pericardial Bioprosthesis: Clinical and Hemodynamic Results
- 1 September 1988
- journal article
- Published by Hindawi Limited in Journal of Cardiac Surgery
- Vol. 3 (3S) , 405-412
- https://doi.org/10.1111/jocs.1988.3.3s.405
Abstract
From 1981 to 1987, 189 patients underwent isolated aortic valve replacement with the Carpentier-Edwards bovine pericardial bioprosthesis. There were 9 early deaths (30-day mortality of 4.8%) and 18 late deaths (3.3% pt-yr). After 5 years, the actuarial survival rate was 81.9%. All patients but one were followed for an average of 37 months after operation, and follow-up totaled 551 patient-years. Valve-related complications inciuded 12 thromboembolic episodes (2.2%/pt-yr), 7 endocarditis (1.3%/pt-yr), 2 hemorrhages due to anticoagulation (0.4%/pt-yr), and 3 reoperations (0.5%/pt-yr). After five years, freedom from thromboembolic events was 91%, that from reoperation was 98%, and 87% for all valve-related complications. There was no primary tissue failure of the valve. Before operation, 55% of the patients were in functional Class III or IV, 97% of the survivors were in Class I or II after valve replacement. Hematological evalvation of 77 patients, an average of 18 months postoperatively, showed no clinically significant hemolysis. Postoperative hemodynamic studies in 28 patients showed mean aortic pressure gradients across the prosthesis at rest ranging from 24 mmHg with the smaller (19 mm) to 11 mmHg with the larger (25 mm and more) valves, and effective orifice areas ranging from 1.0 to 1.9 cm2, respectively. Thus, the Carpentier-Edwards pericardial bioprosthesis appears to be a reliable and hemodynamically effective substitute for replacement of the aortic valve.Keywords
This publication has 13 references indexed in Scilit:
- Thrombotic and Bleeding Complications of Prosthetic Heart ValvesThe Annals of Thoracic Surgery, 1987
- Six- to ten-year follow-up of patients with the Hancock cardiac bioprosthesisThe Journal of Thoracic and Cardiovascular Surgery, 1986
- Comparative study of primary tissue valve failure between lonescu-shiley pericardial and Hancock porcine valves in the aortic positionThe American Journal of Cardiology, 1986
- Rest and exercise hemodynamics following aortic valve replacementThe Journal of Thoracic and Cardiovascular Surgery, 1985
- Incidence of primary tissue valve failure with the Ionescu-Shiley pericardial valveThe Journal of Thoracic and Cardiovascular Surgery, 1985
- The porcine bioprosthetic valveThe Journal of Thoracic and Cardiovascular Surgery, 1985
- Ionescu-Shiley Pericardial Xenografts: Follow-up of Up to 6 YearsThe Annals of Thoracic Surgery, 1985
- Valve failure with the Ionescu-Shiley bovine pericardial bioprosthesis: Analysis of 2680 patientsJournal of Vascular Surgery, 1985
- Early and late risk of aortic valve replacementThe Journal of Thoracic and Cardiovascular Surgery, 1984
- Five-Year Experience with the Ionescu-Shiley Bovine Pericardial Valve in the Aortic PositionThe Annals of Thoracic Surgery, 1983