Flow Rate-Pressure Drop Relation in Coronary Angioplasty: Catheter Obstruction Effect
- 1 February 1996
- journal article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 118 (1) , 83-89
- https://doi.org/10.1115/1.2795949
Abstract
Quantitative methods to measure the hemodynamic consequences of various endovascular interventions including balloon angioplasty are limited. Catheters measuring translesional pressure drops during balloon angioplasty procedures can cause flow blockage and thus inaccurate estimates of pre- and post-intervention flow rates. The purpose of this investigation was to examine the influence of the presence and size of an angioplasty catheter on measured mean pressure gradients across human coronary artery stenoses. Analytical flow modeling and in vitro experimental evidence, coupled with angiographic data on the dimensions and shape of stenotic vessel segments before and after angioplasty, indicated significant flow blockage effects with the catheter present.Keywords
This publication has 17 references indexed in Scilit:
- Estimated mean flow resistance increase during coronary artery catheterizationJournal of Biomechanics, 1994
- Intravascular Ultrasound Imaging: The New Standard for Guidance and Assessment of Endovascular Interventions?Journal of Clinical Laser Medicine & Surgery, 1992
- Flow-dependent coronary artery dilatation in humans.Circulation, 1989
- Measurement of transstenotic pressure gradient during percutaneous transluminal coronary angioplasty.Circulation, 1986
- Usefulness of transstenotic coronary pressure gradient measurements during diagnostic catheterizationThe American Journal of Cardiology, 1985
- Phasic pressure gradients across coronary and renal artery stenoses in humansAmerican Heart Journal, 1983
- Experimental study of pulsatile and steady flow through a smooth tube and an atherosclerotic coronary artery casting of manJournal of Biomechanics, 1983
- Nonoperative Dilatation of Coronary-Artery StenosisNew England Journal of Medicine, 1979
- The pulsed Doppler coronary artery catheter preliminary report of a new technique for measuring rapid changes in coronary artery flow velocity in man.Circulation, 1977
- Results of laminar flow analysis and turbulent flow experiments for eccentric annular ductsAIChE Journal, 1965