Isolation and characterization of wear particles generated in patients who have had failure of a hip arthroplasty without cement.
- 1 September 1995
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 77 (9) , 1301-1310
- https://doi.org/10.2106/00004623-199509000-00002
Abstract
Wear particles from thirty-five membranes obtained during revision hip-replacement operations were studied after digestion of the soft tissue with papain. The particles were isolated and were characterized with use of light and scanning electron microscopic techniques, x-ray microanalysis, and an automated particle analyzer. The mean size of the polyethylene particles was 0.5 micrometer, and the metal particles were a mean of 0.7 micrometer, as determined with scanning electron microscopy. The automated particle analyzer revealed a mean particle diameter of 0.63 micrometer (more than 90 per cent of all particles were less than 0.95 micrometer) and a mean of 1.7 billion particles per gram of tissue, compared with only 143 million per gram of tissue for the control samples. X-ray microanalysis revealed metal debris in sixteen (46 per cent) of the thirty-five membranes after digestion. Thirteen (50 per cent) of the twenty-six membranes surrounding a titanium-alloy stem contained metal particles, compared with three of the nine membranes surrounding a chromium-cobalt stem. Metal debris was present in only one of the twelve membranes surrounding a titanium-alloy stem without a porous coating, compared with twelve of the fourteen membranes surrounding a titanium-alloy stem with a porous coating. This tenfold difference in prevalence was significant (p < 0.005). On the average, the total number of particles (expressed in millions per gram of tissue) associated with the bipolar acetabular components was twice that associated with the fixed acetabular components. In addition, there was a trend toward a larger mean size of the polyethylene particles in association with the bipolar cups. Our data indicate that particulate prosthetic debris in the tissues around failed femoral components that have been inserted without cement constitutes a class of particles that are predominantly less than one micrometer in size and are present in amounts of more than one billion particles per gram of tissue. Routine histological methods did not detect this class of wear debris and led to a gross underestimation of the amount of debris in these membranes.Keywords
This publication has 36 references indexed in Scilit:
- Severe osteolysis of the pelvic in association with acetabular replacement without cement.Journal of Bone and Joint Surgery, 1993
- Bone resorption activity of particulate-stimulated macrophagesJournal of Bone and Mineral Research, 1993
- Size of metallic and polyethylene debris particles in failed cemented total hip replacementsThe Journal of Bone and Joint Surgery. British volume, 1992
- Macrophage exposure to polymethyl methacrylate leads to mediator release and injuryJournal of Orthopaedic Research, 1991
- Bone lysis in well-fixed cemented femoral componentsThe Journal of Bone and Joint Surgery. British volume, 1990
- The synovial production of collagenase and chondrocyte activating factors in response to cobaltJournal of Orthopaedic Research, 1988
- Metallic wear in failed titanium-alloy total hip replacements. A histological and quantitative analysis.Journal of Bone and Joint Surgery, 1988
- Ion implantation of surgical Ti‐6Al‐4V for improved resistance to wear‐accelerated corrosionJournal of Biomedical Materials Research, 1987
- The biologic interface between bone and cementless femoral endoprosthesesThe Journal of Arthroplasty, 1986
- Rate of Wear in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975