Progress in the fabrication of γ-ray detectors from high purity germanium
- 1 February 1971
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 18 (1) , 166-169
- https://doi.org/10.1109/TNS.1971.4325858
Abstract
A planar detector with a depletion layer width of 8mm has been fabricated from high purity germanium. A total system resolution of 2.4 keV at 1.33MeV is obtained with 600 volts bias on this detector. Another detector with a 3mm thick depletion region which was operated at 2000 volts bias yielded a system resolution of 1.65 keV at 1.33 MeV. Thin window detectors have been fabricated by using a glow discharge of BF3 or PF5. This technique has also been used to form ohmic contacts on high purity germanium.Keywords
This publication has 2 references indexed in Scilit:
- Gamma Ray Detectors Made from High Purity GermaniumIEEE Transactions on Nuclear Science, 1970