An EPROM‐based programmable contour generator for use in flow cytometry
- 1 September 1988
- Vol. 9 (5) , 491-493
- https://doi.org/10.1002/cyto.990090514
Abstract
An erasable programmable read‐only memory (EPROM) contour generator has been fabricated to produce contours for use in flow cytometry. Contours are analog waveforms representing the fluorescence or light‐scatter intensity distribution along a cell or object. The generator has particular utility in the development and testing of slit‐scan instrumentation and analysis alogrithms. Contours are generated without the requirement of specimens or full operation of the flow instrumentation. The generator provides control of contour height, width, offset, and rate. The EPROM may be custom programmed to produce contours for specific test applications or for reproducing “real” contour events. The generator is useful in situations where constant repetitive contours of predetermined characteristics are required.Keywords
This publication has 2 references indexed in Scilit:
- A Two-Station Multidimensional Slit-Scan Flow System: Concept and Optical ImplementationIEEE Transactions on Biomedical Engineering, 1982
- A multidimensional slit-scan flow system.Journal of Histochemistry & Cytochemistry, 1979