THE CARDIAC OUTPUT IN MAN: STUDIES WITH THE LOW FREQUENCY, CRITICALLY-DAMPED BALLISTOCARDIOGRAPH, AND THE METHOD OF RIGHT ATRIAL CATHETERIZATION 1
Open Access
- 1 January 1947
- journal article
- research article
- Published by American Society for Clinical Investigation in Journal of Clinical Investigation
- Vol. 26 (1) , 1-10
- https://doi.org/10.1172/jci101780
Abstract
The details of the construction of this ballistocardiograph are presented and the procedures for its adjustment to definite physical conditions described. The adjustments enable the reproduction of this type of ballistocardiograph to be made easily, and permit, without further calibration by the catheter method, the use of the stroke vol. equation tested in this study. Eighty-one detns. of the cardiac output by the method of right atrial catheterization and by almost simultaneous ballistic records with the ballistocardiograph described here have been made on a group of 08 normal subjects and hospital patients under a variety of conditions. Of the 54 detns. on subjects without heart disease, 87% of the ballistic values were within 25% of the catheter values. For 10 patients with aortic insufficiency, the ballistic values were generally much larger than the catheter values. For 17 other cardiac patients the ballistic values were within 25% of the catheter values in 60% of the cases. Serial measurements on patients in shock and during recovery showed a good correlation between the cardiac output measurements by the 2 methods. This observation suggests that changes in cardiac output in any individual subject may be measured with considerable accuracy by means of the low frequency, critically damped ballistocardiograph. It is concluded that this ballistocardiograph provides a useful and reasonably accurate method of determining cardiac output in normal subjects, and in patients with a variety of disease conditions.Keywords
This publication has 5 references indexed in Scilit:
- THE CARDIAC OUTPUT IN MAN: A STUDY OF SOME OF THE ERRORS IN THE METHOD OF RIGHT HEART CATHETERIZATIONAmerican Journal of Physiology-Legacy Content, 1946
- THE CARDIAC OUTPUT IN MALE SUBJECTS AS MEASURED BY THE TECHNIQUE OF RIGHT ATRIAL CATHETERIZATION. NORMAL VALUES WITH OBSERVATIONS ON THE EFFECT OF ANXIETY AND TILTING 1Journal of Clinical Investigation, 1945
- THE DESIGN OF THE BALLISTOCARDIOGRAPHAmerican Journal of Physiology-Legacy Content, 1944
- COMPARISON OF RESULTS OF THE NORMAL BALLISTOCARDIOGRAM AND A DIRECT FICK METHOD IN MEASURING THE CARDIAC OUTPUT IN MAN 1Journal of Clinical Investigation, 1942
- STUDIES ON THE ESTIMATION OF CARDIAC OUPTUT IN MAN, AND OF ABNORMALITIES IN CARDIAC FUNCTION, FROM THE HEART'S RECOIL AND THE BLOOD'S IMPACTS; THE BALLISTOCARDIOGRAMAmerican Journal of Physiology-Legacy Content, 1939