Optimization of metal/phosphor screens for on-line portal imaging
- 1 February 1994
- journal article
- Published by Wiley in Medical Physics
- Vol. 21 (2) , 227-235
- https://doi.org/10.1118/1.597299
Abstract
Studies were conducted to determine the optimal metal/phosphor screen for on-line video verification of radiation treatment portals. Screens were evaluated for luminance and spatial resolution as a function of composition and thickness at 6- and 23-MV x-ray energies. A new video technique was used to determine modulation transfer functions. Gd2O2S was found to be the most efficient (brightest) phosphor for this application. Luminance was found to vary linearly with phosphor thickness up to a thickness of 500 mg/cm2. Metal plates made of iron, brass, copper, lead, and sintered tungsten of various thicknesses were also tested for luminance and resolution with Gd2O2S phosphor. Brightness peaked at about 2-mm thickness for most metals. Significant contributions to the brightness were found to come from x rays interacting with the phosphor itself.Keywords
This publication has 0 references indexed in Scilit: