Death Due to an Implantable Cardioverter Defibrillator
- 5 August 2004
- journal article
- case report
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 15 (8) , 953-956
- https://doi.org/10.1046/j.1540-8167.2004.03611.x
Abstract
Inappropriate therapy due to noise oversensing caused a true ventricular fibrillation (VF) and death of a patient. A 49-year-old patient with a history of dilated cardiomyopathy received a double-chamber implantable cardioverter defibrillator (ICD) in 1991 for a sustained inducible ventricular tachycardia (VT). One appropriate shock delivered in 1994 terminated an episode of VT. The generator was replaced in 1995 and in 2000, and was connected to the initial leads. Three months after the second replacement, the patient received six consecutive shocks related to detection of noise interpreted as VF. Unfortunately, the sixth shock triggered a true VF, which was not treated due to end of the therapeutic sequence, and the patient died. The causes of the dysfunction are discussed.Keywords
This publication has 19 references indexed in Scilit:
- Detection and management of animplantable cardioverter defibrillator lead failureJournal of the American College of Cardiology, 2003
- Early Detection of Lead Fracture by Painless High Voltage Lead Impedance Measurement in a Transvenous ICD Lead SystemJournal of Interventional Cardiac Electrophysiology, 2000
- Adverse Events With Transvenous Implantable Cardioverter-DefibrillatorsCirculation, 1998
- Sensing Lead-Related Complications in Patients With Transvenous Implantable Cardioverter-DefibrillatorsThe American Journal of Cardiology, 1996
- The Proarrhythmic Potential of Implantable Cardioverter-DefibrillatorsCirculation, 1995
- Limitations and Late Complications of Third-Generation Automatic Cardioverter-DefibrillatorsCirculation, 1995
- Oversensing during ventricular pacing in patients with a third-generation implantable cardioverter-defibrillatorJournal of the American College of Cardiology, 1994
- Efficacy of automatic multimodal device therapy for ventricular tachyarrhythmias as delivered by a new implantable pacing cardioverter-defibrillator. Results of a European multicenter study of 102 implants.Circulation, 1992
- Holter Documented Sudden Death in a Patient with an Implanted DefibrillatorPacing and Clinical Electrophysiology, 1992
- A hemodynamically responsive antitachycardia system. Development and basis for design in humans.Circulation, 1990