Assay of Uranium in the Presence of Various Interfering Ions Using Ion Pair Liquid Chromatography
- 1 May 1993
- journal article
- research article
- Published by Taylor & Francis in Journal of Liquid Chromatography
- Vol. 16 (7) , 1589-1599
- https://doi.org/10.1080/10826079308020975
Abstract
A sensitive and specific method for uranium(VI) determination using reverse phase ion-pair high performance liquid chromatography with UV detector has been developed. The developed method has been applied for the determination of uranium in a synthetic sample containing most of the interfering ions. Except iron most of the elements and anion which often interfere in uranium assay do not interfere in the present method. Interference of iron is masked by addition of 5mM/l EDTA solution to the sample.Keywords
This publication has 6 references indexed in Scilit:
- Determination of uranium in technological waters by ion-pair liquid chromatographyTalanta, 1991
- Development of a Rapid, Economical and Sensitive Method for the Routine Determination of Excreted Uranium in UrineAnalytical Letters, 1991
- Determination of uranium in natural groundwaters using high-performance liquid chromatographyAnalytical Chemistry, 1988
- Continuous flow solvent extraction system for the determination of trace amounts of uranium in nuclear waste reprocessing solutionsThe Analyst, 1988
- Determination of uranium(VI) in process liquors by ion chromatographyThe Analyst, 1987
- Spectrophotometric determination of uranium(vi) with 2-(5-bromo-2-pyridylazo)-5-diethylaminophenolAnalytica Chimica Acta, 1971