Die Röntgendigitalanlage als Instrument zur kinematischen Untersuchung der Schulter — Eine Untersuchung zur Strahlenbelastung - Digital Radiography as a Tool for the Kinematic Analysis of the Shoulder Joint - A Study of Radiation Dose
- 1 January 1991
- journal article
- research article
- Published by Walter de Gruyter GmbH in Biomedizinische Technik/Biomedical Engineering
- Vol. 36 (6) , 149-152
- https://doi.org/10.1515/bmte.1991.36.6.149
Abstract
The radiation exposure during biomechanical investigation of the shoulder joint done with a digital radiography system (Polytron 1000 VR - Siemens) was measured experimentally. Total radiation dose and surface dose were determined by phantom measurements. The radiation exposure per image is about 1/6 to 1/7 of that of a conventional study. The digital radiography system used is thus suitable for kinesiologic studies of large joints in healthy volunteers.Keywords
This publication has 1 reference indexed in Scilit:
- DIE KINEMATISCHE ANALYSE DER SCHULTERABDUKTION UNTER DYNAMISCHEN BEDINGUNGEN. Eine Untersuchung mit der RöntgenanlageBiomedizinische Technik/Biomedical Engineering, 1989