Abstract
Noise power spectrum (NPS) and detective quantum efficiency (DQE) for metal-plate/film portal detectors are reported for the Co-60 and Linac 10 MV spectra. The detectors consist of a double-emulsion portal film secured between plates of aluminum, copper, brass, or lead. The NPS was found to be independent of the detector sensitivity demonstrating that the film grain noise dominates over the quantum noise. Although the detector signal increases with density of the backplates, the resultant electron-backscatter increases detector blur, thus decreasing the DQE. The lowest DQEs are therefore produced with the lead backplates, irrespective of the front plates. For a given backplate and for front-plate thicknesses less than the maximum electron range, the DQE increases with density of the front plate.Keywords
This publication has 7 references indexed in Scilit:
- Characteristics of metal‐plate/film detectors at therapy energies. I. Modulation transfer functionMedical Physics, 1998
- Therapy imaging: A signal-to-noise analysis of metal plate/film detectorsMedical Physics, 1987
- Signal-to-noise properties of mammographic film-screen systemsMedical Physics, 1985
- Fluence to Dose Equivalent Conversion Factors Calculated With EGS3 for Electrons from 100 keV to 20 GeV and Photons from 11 keV to 20 GeVHealth Physics, 1984
- Fast Fourier digital quantum mottle analysis with application to rare earth intensifying screen systemsMedical Physics, 1977
- Noise Measurements On Rare-Earth Intensifying Screen SystemsPublished by SPIE-Intl Soc Optical Eng ,1974
- A Unified Approach to the Performance of Photographic Film, Television Pickup Tubes, and the Human EyeJournal of the Society of Motion Picture Engineers, 1946