Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters
Top Cited Papers
- 21 February 2002
- journal article
- Published by Wiley in Medical Physics
- Vol. 29 (3) , 379-390
- https://doi.org/10.1118/1.1446109
Abstract
The BEAM code is used to simulate nine photon beams from three major manufacturers of medical linear accelerators (Varian, Elekta, and Siemens), to derive and evaluate estimates for the parameters of the electron beam incident on the target, and to study the effects of some mechanical parameters like target width, primary collimator opening, flattening filter material and density. The mean energy and the FWHM of the incident electron beam intensity distributions (assumed Gaussian and cylindrically symmetric) are derived by matching calculated percentage depth‐dose curves past the depth of maximum dose (within 1% of maximum dose) and off‐axis factors (within 2σ at 1% statistics or less) with measured data from the AAPM RTC TG‐46 compilation. The off‐axis factors are found to be very sensitive to the mean energy of the electron beam, the FWHM of its intensity distribution, its angle of incidence, the dimensions of the upper opening of the primary collimator, the material of the flattening filter and its density. The off‐axis factors are relatively insensitive to the FWHM of the electron beam energy distribution, its divergence and the lateral dimensions of the target. The depth‐dose curves are sensitive to the electron beam energy, and to its energy distribution, but they show no sensitivity to the FWHM of the electron beam intensity distribution. The electron beam incident energy can be estimated within 0.2 MeV when matching either the measured off‐axis factors or the central‐axis depth‐dose curves when the calculated uncertainties are about 0.7% at the 1 σ level. The derived FWHM (±0.1 mm) of the electron beam intensity distributions all fall within 1 mm of the manufacturer specifications except in one case where the difference is 1.2 mm.Keywords
This publication has 38 references indexed in Scilit:
- Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantomMedical Physics, 2001
- Comparison of measured and Monte Carlo calculated dose distributions from the NRC linacMedical Physics, 2000
- Performance benchmarks of the MCV Monte Carlo systemPublished by Springer Nature ,2000
- OMEGA Monte Carlo verification of dosimetric data for Gamma Knife®Published by Springer Nature ,2000
- Clinical implementation of a Monte Carlo treatment planning systemMedical Physics, 1999
- A New Class of Random Number GeneratorsThe Annals of Applied Probability, 1991
- Characteristics of photon beams from Philips SL25 linear acceleratorsMedical Physics, 1990
- Dosimetric characteristics of a 6 MV photon beam from a linear accelerator with asymmetric collimator jawsInternational Journal of Radiation Oncology*Biology*Physics, 1988
- The effect of off-axis quality changes on zero area TAR for megavoltage beamsPhysics in Medicine & Biology, 1985