Evaluation of matrix effects in ion microscopic analysis of freeze‐fractured, freeze‐dried cultured cells
- 1 December 1987
- journal article
- research article
- Published by Wiley in Journal of Microscopy
- Vol. 148 (3) , 223-239
- https://doi.org/10.1111/j.1365-2818.1987.tb02869.x
Abstract
SIMS matrix effects (mass interferences, sputter yield variations and practical ion yield variations) were evaluated in freeze-fractured, freeze-dried cultured cells at the ∼0.5 μm spatial resolution of the Cameca IMS-3f ion microscope. Cell lines studied include normal rat kidney (NRK), 3T3 mouse fibroblast, L6 rat myoblast, chinese hamster ovary (CHO) and rat kangaroo kidney (PtK2) cells. High mass resolution studies indicated that the secondary ion signals of H—, C—, O—, Na+, Mg+, CN—, P—, S—, Cl—, K+ and Ca+ were free from major mass interferences. However, a large mass interference was observed for nitrogen at mass 14. No significant sputtering yield difference between the nuclear and cytoplasmic compartments of the cells studied was observed. The subcellular distributions of the major (H, C, N and O) and minor (P, S, K, Cl, Na, Mg and Ca) matrix elements were found to be largely homogeneous with the exception of Ca, which was observed mainly in the cell cytoplasm. Practical ion yield variations were compared by three different approaches: (i) by the use of cells doped with known electrolyte concentrations, (ii) by quantitative ion implantation, and (iii) by analysis of the same cell with both electron probe and ion microscope. Each approach indicated an absence of significant practical ion yield differences between the nuclear and cytoplasmic regions of these specimens. These observations indicate that secondary ion signals in this type of sample are not significantly affected by local matrix effect variations. Hence, qualitative imaging of such specimens provides a true representation of subcellular elemental distribtions. These observations should allow the development of quantitative ion imaging methodologies and enhance the applicability of ion microscopy to biomedical problems.Keywords
This publication has 22 references indexed in Scilit:
- A low-viscosity epoxy resin embedding medium for electron microscopyPublished by Elsevier ,2004
- Ionization probability variations due to matrix in ion microscopic analysis of plastic-embedded and ashed biological specimensAnalytical Chemistry, 1986
- Intracellular localization of diffusible elements in frozen-hydrated biological specimens with ion microscopyAnalytical Chemistry, 1986
- Imaging Elemental Distribution and Ion Transport in Cultured Cells with Ion MicroscopyScience, 1985
- Relative sensitivity factors of elements in quantitative secondary ion mass spectrometric analysis of biological reference materialsAnalytical Chemistry, 1983
- Ion implantation for quantitative ion microscopy of biological soft tissueAnalytical Chemistry, 1983
- Applications of secondary ion mass spectrometry (SIMS) in biological research: a reviewJournal of Microscopy, 1982
- PREFERENTIAL SPUTTERING EFFECTS IN THE ION MICROANALYSIS OF BIOLOGICAL TISSUESJournal of Microscopy, 1981
- Applications of SIMS in biology and medicineScanning, 1980
- Ion microscopyAnalytical Chemistry, 1975