Bubbling Humidifiers Produce Microaerosols Which Can Carry Bacteria
- 1 June 1986
- journal article
- research article
- Published by Cambridge University Press (CUP) in Infection Control
- Vol. 7 (8) , 403-407
- https://doi.org/10.1017/s0195941700064638
Abstract
It is widely held that bubbling humidifiers do not produce microaerosols, although prior studies have resulted in conflicting evidence. We have studied this phenomenon in a clean room using an airborne particle counter and samplers for airborne bacteria. At gas flow rates between 10 and 80 L/min, a Cascade 1 humidifier produced between 460 and 999 water droplets/L humidified gas. Total water volume aerosolized was approximately 10-8 ml/L humidified gas. Seventy-three percent of the particles had diameters between 1 and 5 microns. With the reservoir containing 6.4 × 106P. aeruginosa/ml, it produced between 2 and 9 P. aeruginosa/Lhumidified gas. Most of the bacteria were in particles of a size likely to be deposited in the lung. This bacterial carryover was between 20 and 150 times the amount predicted by multiplication of the water volume aerosolized times the concentration of bacteria in the humidifier reservoir. An Air Life humidifier produced fewer particles which were also of a size likely to be deposited in the lung and, when the reservoir contained P. aeruginosa, it aerosolized bacteria. Wick-type humidifiers did not produce detectable aerosol or bacterial carry-over. Although the clinical significance of these findings has not been established, they provide a rationale for the CDC recommendations for procedures designed to keep bubbling humidifier reservoir water uncontaminated.This publication has 14 references indexed in Scilit:
- Surveillance study of pediatric patients receiving respiratory therapy, using the bain circuit systemAmerican Journal of Infection Control, 1982
- Contamination of Mechanical Ventilators with Tubing Changes Every 24 or 48 HoursNew England Journal of Medicine, 1982
- Guideline for Prevention of Nosocomial PneumoniaInfection Control, 1982
- A method of maintaining sterility of heated humidifiers in mechanical ventilator breathing circuitsCritical Care Medicine, 1978
- GENERATION OF AEROSOLS BY BURSTING OF SINGLE BUBBLESPublished by Elsevier ,1976
- An Evaluation of Bacterial Contamination of Ventilator Humidifying SystemsChest, 1973
- Serratia marcescens Infections from Inhalation Therapy Medications: Nosocomial OutbreakAnnals of Internal Medicine, 1970
- The Potential Role of Inhalation Therapy Equipment in Nosocomial Pulmonary Infection*Journal of Clinical Investigation, 1965
- Separation of Microorganisms by FlotationApplied Microbiology, 1960
- Separation of Microorganisms by FlotationApplied Microbiology, 1960