Haemodynamic benefit of a rate-adapted A-V delay in dual chamber pacing
- 30 June 1989
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 10 (7) , 637-646
- https://doi.org/10.1093/oxfordjournals.eurheartj.a059541
Abstract
In dual chamber pacing, an improvement of exercise capacity is expected when the atrial refractory period is shortened, because the 2/1 point is increased. This objective can be achieved by greatly reducing atrio-ventricular delay (AVD) on exercise. Are such variations (up to 100–120 ms) detrimental from a haemodynamic standpoint? This study was performed to analyse this particular aspect of DDD pacing. Three DDD pacing modes, differing by their AVDs (fixed 200 ms AVD, fixed 150 ms AVD, and rate-adapted AVD) were tested in random order, with a haemodynamic protocol including ten patients with chronic atrio-ventricular (A-V) block. For the rate-adapted AVD pacing mode, AVD was reduced by 20 ms every 10beats min−1 increment (from 220 ms at 90 beats min−1 to 100 ms at 150 beats min−1). Pacing rate was increased from 90 to 150 beats min−1 by increments of 10 beats min−1 every 5 min. Cardiac performance was significantly improved with the rate-adapted AVD above the two fixed AVDs, despite a large A VD variation. When A VD was rate adapted, cardiac index, stroke volume index and left ventricular systolic work index were generally higher and pulmonary capillary wedge pressure, pulmonary arterial pressure and systemic vascular resistances were generally lower, especially at 120, 130 and 140 beats min−1. Comparing the two fixed AVDs, 200 AVD improved cardiac function more at lower heart rates, whereas 150 AVD improved cardiac function more at higher heart rates. Despite its limitations, this study demonstrates that the potential benefits of reducing AVD with increasing heart rates should be twofold in dual chamber pacing:Keywords
This publication has 10 references indexed in Scilit:
- Importance of left atrial timing in the programming of dual-chamber pacemakersThe American Journal of Cardiology, 1987
- Left ventricular function at similar heart rates during tachycardia induced by exercise and atrial pacing: an echocardiographic study.Heart, 1987
- Evaluation by pulsed doppler echocardiography of the atrial contribution to left ventricular filling in patients with DDD pacemakersThe American Journal of Cardiology, 1986
- Hemodynamic comparison of ventricular pacing, atrioventricular sequential pacing, and atrial synchronous ventricular pacing using radionuclide ventriculographyThe American Journal of Cardiology, 1986
- The hemodynamic consequences of cardiac arrhythmias: Evaluation of the relative roles of abnormal atrioventricular sequencing, irregularity of ventricular rhythm and atrial fibrillation in a canine modelAmerican Heart Journal, 1983
- A comparison of the acute and long-term hemodynamic effects of ventricular inhibited and atrial synchronous ventricular inhibited pacing.Circulation, 1982
- Non‐Invasive Beat‐to‐Beat Analysis of Stroke Volume and Digital Pulse Volume in Patients with Complete Heart Block and Artificial PacingActa Medica Scandinavica, 1979
- P-R interval in relation to heart rate during exercise and the influence of posture and autonomic toneJournal of Electrocardiology, 1977
- Hemodynamic consequences of sequential atrioventricular pacingThe American Journal of Cardiology, 1968
- The relationship between the timing of atrial systole and the useful work of the left ventricle in manAmerican Heart Journal, 1966