A System for Mounting End Caps on Ice Specimens

Abstract
This short note describes the equipment and procedures we have developed to mount end caps on ice-core specimens. The system typically achieves end-plane parallelism within 0.5 μm/mm of specimen diameter (i.e. a total indicator run-out of 0.002 in for a 4.0 in diameter specimen). The essential elements of the system are a holder and an alignment fixture. The holder firmly grips the ice core about its circumference by the compression of two series of O–rings. The alignment fixture clamps the holder to align the ice core precisely with the end caps. To bond the ice to the end cap we form a layer of 0°C water on the end cap; the water freezes immediately upon contact with the ice and forms a strong intimate bond. To date, we have used this system to install phenolic end caps on 101.6 mm diameter cores and aluminum end caps on 76.2 mm diameter cores of saline ice. We obtained a somewhat better tolerance with the aluminum caps (0.33 μm/mm of end-cap diameter), due primarily to the geometric stability of that material under the prevailing conditions. These specimens have been successfully tested in uniaxial and triaxial compression, and we expect that with appropriate end caps the system will be suitable for preparing tension specimens as well.