Comparative Evaluation of the Performance of the Abbott Real-Time Human Immunodeficiency Virus Type 1 (HIV-1) Assay for Measurement of HIV-1 Plasma Viral Load following Automated Specimen Preparation

Abstract
The performance of the new Abbott real-time human immunodeficiency virus type 1 (HIV-1) assay for HIV-1 RNA load determination in plasma was compared to that of the Abbott LCx HIV-1 RNA quantitative assay following automated RNA isolation by the Abbott m1000 extractor. The measured viral loads of 89 clinical specimens differed by mean 0.19 log 10 copies/ml (95% confidence interval, 0.12 to 0.26 log 10 copies/ml). Although the difference in viral load determinations was positively skewed in favor of the LCx assay, it did not reach statistical significance ( P = 0.42). Results were linearly associated ( R 2 = 0.94) and strongly correlated ( R = 0.96). Good performance was observed with HIV-1 subtypes other than B and circulating recombinant forms, although results obtained with two subtype G specimens and one H specimen showed a more substantial difference.