Microfluidic Shear Devices for Quantitative Analysis of Cell Adhesion
Top Cited Papers
- 17 August 2004
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 76 (18) , 5257-5264
- https://doi.org/10.1021/ac049837t
Abstract
We describe the design, construction, and characterization of microfluidic devices for studying cell adhesion and cell mechanics. The method offers multiple advantages over previous approaches, including a wide range of distractive forces, high-throughput performance, simplicity in experimental setup and control, and potential for integration with other microanalytic modules. By manipulating the geometry and surface chemistry of the microdevices, we are able to vary the shear force and the biochemistry during an experiment. The dynamics of cell detachment under different conditions can be captured simultaneously using time-lapse videomicroscopy. We demonstrate assessment of cell adhesion to fibronectin-coated substrates as a function of the shear stress or fibronectin concentration in microchannels. Furthermore, a combined perfusion-shear device is designed to maintain cell viability for long-term culture as well as to introduce exogenous reagents for biochemical studies of cell adhesion regulation. In agreement with established literature, we show that fibroblasts cultured in the combined device reduced their adhesion strength to the substrate in response to epidermal growth factor stimulation.Keywords
This publication has 16 references indexed in Scilit:
- Biophysical Integration of Effects of Epidermal Growth Factor and Fibronectin on Fibroblast MigrationBiophysical Journal, 1999
- Short-Term Cell/Substrate Contact Dynamics of Subconfluent Endothelial Cells following Exposure to Laminar FlowBiotechnology Progress, 1999
- Microfluidic Diffusion-Based Separation and DetectionScience, 1999
- Rapid Prototyping of Microfluidic Systems in Poly(dimethylsiloxane)Analytical Chemistry, 1998
- A Theoretical Model Study of the Influence of Fluid Stresses on a Cell Adhering to a Microchannel WallBiophysical Journal, 1998
- Cell movement elicited by epidermal growth factor receptor requires kinase and autophosphorylation but is separable from mitogenesisThe Journal of cell biology, 1994
- A numerical analysis of forces exerted by laminar flow on spreading cells in a parallel plate flow chamber assayBiotechnology & Bioengineering, 1993
- Design and construction of a linear shear stress flow chamberAnnals of Biomedical Engineering, 1993
- Micromanipulation of adhesion of a Jurkat cell to a planar bilayer membrane containing lymphocyte function-associated antigen 3 molecules.The Journal of cell biology, 1992
- The Effect of Fluid Mechanical Stress on Cellular Arachidonic Acid MetabolismaAnnals of the New York Academy of Sciences, 1987