A rapid 2D time‐resolved variable‐rate k‐space sampling MR technique for passive catheter tracking during endovascular procedures
- 1 September 1998
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 40 (3) , 356-362
- https://doi.org/10.1002/mrm.1910400304
Abstract
A new, fast, 2D MR imaging technique allowing passive catheter visualization adequate for use as a tool for guiding the movement of a catheter during endovascular procedures is described. This imaging technique samples low spatial frequencies more often than high spatial frequencies; it also uses both k-space view sharing and temporal interpolation. Unlike other techniques for passive visualization that exploit magnetic-susceptibility–induced artifacts, we have adopted a strategy that takes advantage of the T1-shortening effect of paramagnetic contrast agents, such as Gd-DTPA and a projection dephaser. This not only permits visualization of the entire catheter length but also minimizes the risk of intravascular heating. Using this method, a temporal frame rate of up to eight images per second and a tip localization accuracy of ± 1mm (root mean square difference) can be achieved.Keywords
This publication has 5 references indexed in Scilit:
- Application of keyhole imaging to interventional MRI: A simulation study to predict sequence requirementsJournal of Magnetic Resonance Imaging, 1996
- Needle localization in MR-guided biopsy and aspiration: effects of field strength, sequence design, and magnetic field orientation.American Journal of Roentgenology, 1996
- Reduction of partial‐volume artifacts with zero‐filled interpolation in three‐dimensional MR angiographyJournal of Magnetic Resonance Imaging, 1994
- “Keyhole” method for accelerating imaging of contrast agent uptakeJournal of Magnetic Resonance Imaging, 1993
- Real‐time position monitoring of invasive devices using magnetic resonanceMagnetic Resonance in Medicine, 1993