Respiratory mechanics in mechanically ventilated patients with respiratory failure
- 1 June 1985
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 58 (6) , 1849-1858
- https://doi.org/10.1152/jappl.1985.58.6.1849
Abstract
In 11 mechanically ventilated patients, respiratory mechanics were measured 1) during constant flow inflation and 2) following end-inflation airway occlusion, as proposed in model analysis (J. Appl. Physiol. 58: 1840–1848, 1985. During the latter part of inflation, the relationship between driving pressure and lung volume change was linear, allowing determination of static respiratory elastance (Ers) and resistance (RT). The latter represents in each patient the maximum resistance value that can obtain with the prevailing time constant inhomogeneity. Following occlusion, Ers and RT were also obtained along with RT (min) which represents a minimum, i.e., resistance value that would obtain in the absence of time constant inhomogeneity. A discrepancy between inflation and occlusion Ers and RT was found only in the three patients without positive end-expiratory pressure, and could be attributed to recruitment of lung units during inflation. In all instances Ers and RT were higher than normal. RT(min) was lower in all patients than the corresponding values of RT, indicating that resistance was frequency dependent due to time constant inequalities. Changes in inflation rate did not affect Ers, while RT increased with increasing flow.This publication has 12 references indexed in Scilit:
- Single-breath method for measurement of respiratory mechanics in anesthetized animalsJournal of Applied Physiology, 1982
- OCCULT POSITIVE END-EXPIRATORY PRESSURE IN MECHANICALLY VENTILATED PATIENTS WITH AIR-FLOW OBSTRUCTION - THE AUTO-PEEP EFFECTPublished by Elsevier ,1982
- Assessment of airway function during assisted ventilationLung, 1981
- Pulmonary, Chest Wall, and Lung-Thorax Elastances in Acute Respiratory FailureChest, 1981
- A Pulse Method of Measuring Respiratory System Compliance in Ventilated PatientsChest, 1981
- EFFECT OF INSPIRATORY FLOW-RATE ON GAS-EXCHANGE DURING MECHANICAL VENTILATIONPublished by Elsevier ,1981
- A pulse method of measuring respiratory system complianceJournal of Applied Physiology, 1980
- The Mechanics of the Respiratory System During AnesthesiaAnesthesiology, 1965
- Partitioning of respiratory flow resistance in manJournal of Applied Physiology, 1964
- Mechanical Factors in Distribution of Pulmonary VentilationJournal of Applied Physiology, 1956