Reliability of the Dinamap Non-invasive Monitor in the Measurement of Blood Pressure of III Asian Newborns
- 1 May 1990
- journal article
- research article
- Published by SAGE Publications in Clinical Pediatrics
- Vol. 29 (5) , 262-267
- https://doi.org/10.1177/000992289002900502
Abstract
Four hundred thirty-one paired sets of readings of systolic and diastolic blood pressure and 438 paired sets of readings of mean arterial BP from 49 ill newborns, including 21 very low birth weight infants, were analyzed for the extent and pattern of agreement and the linear relationship between the Dinamap oscillometric monitor and the direct intraarterial blood pressure readings. Agreement between the two methods was measured by the intraclass correlation, whereas the linear relationship was assessed by the product-moment correlation. The intraclass correlations for systolic, diastolic and mean blood pressures were 0.696, 0.766, and 0.781, respectively. The product-moment correlations for systolic, diastolic and mean blood pressures were 0.706, 0.768, and 0.786, respectively. BP measurements by the Dinamap monitor showed reasonably close agreement to those obtained by the intraarterial mean arterial pressure ranges above 40 mmHg. For mean arterial pressure of 40 mmHg and lower, BP readings by the Dinamap monitor tended to be higher than those obtained by the intraarterial method. These findings appeared to be consistent regardless of the birth weight of the newborn.Keywords
This publication has 11 references indexed in Scilit:
- Statistical evaluation of agreement between two methods for measuring a quantitative variableComputers in Biology and Medicine, 1989
- Blood pressure in very low birth weight infants in the first 70 days of lifeThe Journal of Pediatrics, 1988
- Blood Pressure in Full-term Healthy NeonatesClinical Pediatrics, 1987
- Dinamap fails to detect hypotension in very low birthweight infants.Archives of Disease in Childhood, 1986
- Blood pressure monitoring in neonates: Comparison of umbilical and peripheral artery catheter measurementsThe Journal of Pediatrics, 1984
- NUTRIENT INTAKES OF HUSBANDS AND WIVES: IMPLICATIONS FOR EPIDEMIOLOGIC RESEARCHAmerican Journal of Epidemiology, 1982
- Clinical evaluation of the Dinamap noninvasive blood pressure monitor in pre-term neonatesClinical Physics and Physiological Measurement, 1981
- Aortic Blood Pressure During the First 12 Hours of Life in Infants with Birth Weight 610 to 4,220 GramsPediatrics, 1981
- An Automated Oscillometric Technique for Estimating Mean Arterial Pressure in Critically III NewbornsAnesthesiology, 1981
- Alternate approaches for quantifying aggregate and individual agreements between two methods for assessing dietary intakesThe American Journal of Clinical Nutrition, 1980