EVALUATION OF 2 TYPES OF MEDICALLY SIGNIFICANT ERROR LIMITS AND 2 QUALITY-CONTROL PROCEDURES ON A MULTICHANNEL ANALYZER
- 1 August 1989
- journal article
- research article
- Vol. 113 (8) , 834-837
Abstract
Analytical error limits based on two different concepts of "medical significance" are derived. The ability of the system to detect medically significant errors using standard (13s) and enchanced (mean and range) quality control procedures is investigated. Error detection is found to be adequate for most analytes when error limits are based on physicians'' opinions and a mean and range quality control system is used. Error detection is unsatisfactory when error limits based on intraindividual biological variation are used and this cannot be improved by using better quality control systems. The limiting factor is analytical precision. The use of error limits based on biological variation as proposed by major clinical chemistry professional bodies cannot be achieved using most current types of analytical tehcnology.This publication has 5 references indexed in Scilit:
- Medically Useful Criteria for Analytic Performance of Laboratory TestsAmerican Journal of Clinical Pathology, 1985
- CLINICAL LABORATORY PRECISION - THE STATE OF THE ART AND MEDICAL USEFULNESS BASED INTERNAL QUALITY-CONTROL1982
- Analytic Precision in Clinical Chemistry and Medical DecisionsAmerican Journal of Clinical Pathology, 1980
- A clinical view of analytical goals in clinical biochemistry.Journal of Clinical Pathology, 1979
- Power functions for statistical control rules.Clinical Chemistry, 1979